Thai Pharmacopoeia · TP-2011-Supplement-2024

Chromatography

TP

Definition

Mobile phase MOBILE PHASE A Trifluoroacetic acid, water, acetonitrile (0.5:950:50) MOBILE PHASE B Trifluoroacetic acid, water, acetonitrile (0.5:50:950) The step gradient of mobile phases is as follows: <table border="1"><tr><td>Time(Minutes)</td><td>Mobile Phase A(Per Cent V/V)</td><td>Mobile Phase B(Per Cent V/V)</td></tr><tr><td>0-8</td><td>97→94</td><td>3→6</td></tr><tr><td>8-25</td><td>94→66</td><td>6→34</td></tr><tr><td>25-40</td><td>66→10</td><td>34→90</td></tr><tr><td>40-45</td><td>10</td><td>90</td></tr></table> ## Chromatographic system Chromatographic system DETECTOR Ultraviolet light (215 nm). COLUMN A stainless steel column (10 cm $ \times $ 2.1 mm), packed with octadecylsilyl bonded to porous silica microparticles, compatible with 100 per cent aqueous mobile phases (5 $ \mu\mathrm{m} $ ) with a pore size of 20 nm, end-capped. COLUMN TEMPERATURE 60° FLOW RATE 0.2 mL per minute System suitability SAMPLE Standard solution Suitability requirements The chromatogram obtained from the standard solution is qualitatively similar to the chromatogram of filgrastim digest supplied with Filgrastim RS. Procedure Inject a portion (about 10 mL) of Standard solution and Test solution into the chromatograph. Measure the areas of the principal peaks in each chromatogram. Identify the peaks due to filgrastim in the test solution using the retention time of filgrastim in the standard solution. Results The profile of the chromatogram obtained from the test solution corresponds to that of the chromatogram obtained from the standard solution. Impurities with molecular weights higher than that of filgrastim Carry out the test as described in the "Size- Exclusion Chromatography" (Appendix 3.6), use the normalization procedure. Diluent Dissolve 4.1 g of sodium acetate in 400 mL of water, adjust to pH 4.0 with acetic acid and dilute to 500 mL with water. Standard solution Dilute Filgrastim RS with Diluent to obtain a concentration of 0.4 mg per mL. Test solution

Monograph text

Preamble

Mobile phase MOBILE PHASE A Trifluoroacetic acid, water, acetonitrile (0.5:950:50) MOBILE PHASE B Trifluoroacetic acid, water, acetonitrile (0.5:50:950) The step gradient of mobile phases is as follows: <table border="1"><tr><td>Time(Minutes)</td><td>Mobile Phase A(Per Cent V/V)</td><td>Mobile Phase B(Per Cent V/V)</td></tr><tr><td>0-8</td><td>97→94</td><td>3→6</td></tr><tr><td>8-25</td><td>94→66</td><td>6→34</td></tr><tr><td>25-40</td><td>66→10</td><td>34→90</td></tr><tr><td>40-45</td><td>10</td><td>90</td></tr></table> Chromatographic system Chromatographic system DETECTOR Ultraviolet light (215 nm). COLUMN A stainless steel column (10 cm \times 2.1 mm), packed with octadecylsilyl bonded to porous silica microparticles, compatible with 100 per cent aqueous mobile phases (5 μm ) with a pore size of 20 nm, end-capped. COLUMN TEMPERATURE 60° FLOW RATE 0.2 mL per minute System suitability SAMPLE Standard solution Suitability requirements The chromatogram obtained from the standard solution is qualitatively similar to the chromatogram of filgrastim digest supplied with Filgrastim RS. Procedure Inject a portion (about 10 mL) of Standard solution and Test solution into the chromatograph. Measure the areas of the principal peaks in each chromatogram. Identify the peaks due to filgrastim in the test solution using the retention time of filgrastim in the standard solution. Results The profile of the chromatogram obtained from the test solution corresponds to that of the chromatogram obtained from the standard solution. Impurities with molecular weights higher than that of filgrastim Carry out the test as described in the "Size- Exclusion Chromatography" (Appendix 3.6), use the normalization procedure. Diluent Dissolve 4.1 g of sodium acetate in 400 mL of water, adjust to pH 4.0 with acetic acid and dilute to 500 mL with water. Standard solution Dilute Filgrastim RS with Diluent to obtain a concentration of 0.4 mg per mL. Test solution Dilute Filgrastim Concentrated Solution with Diluent to obtain a concentration of 0.4 mg per mL. Resolution solution Mix a suitable portion of Standard solution for about 30 seconds using a vortex mixer. Mobile phase Dissolve 7.9 g of ammonium hydrogen carbonate in 1000 mL of water and adjust to pH 7.0 with phosphoric acid; dilute to 2000 mL with water. Chromatographic system DETECTOR Ultraviolet light (215 nm) COLUMN A column (30 cm \times 7.8 mm), packed with spherical silica (5 mm) with hydrophilic coating (suitable for fractionation of globular proteins in the relative molecular mass range of 10,000 to 500,000), equipped with a similarly packed guard column of a suitable size. COLUMN TEMPERATURE 30° FLOW RATE 0.5 mL per minute System suitability SAMPLE Resolution solution (Note The relative retention with reference to the filgrastim monomer: aggregates, filgrastim oligomer 1, filgrastim oligomer 2, and filgrastim dimer are about 0.60,0.75,0.80,and 0.85,respectively,where the retention time for filgrastim monomer is about 19 minutes.) Suitability requirements RETENTION TIME 17 to 20 minutes for filgrastim monomer. RESOLUTION Not less than 3 between filgrastim dimer and filgrastim monomer peaks. Procedure Separately inject equal volumes (about 20 μL ) of Standard solution and Test solution into the chromatograph, record the chromatograms and measure the responses for the major peaks. Calculation Calculate the percentage content of the dimer, oligomers and aggregates.

Limits

- impurities with molecular weights higher than that of filgrastim, other than the dimer: not more than 0.5 per cent; - total of impurities with molecular weights higher than that of filgrastim: not more than 2 per cent. Impurities with molecular weights differing from that of filgrastim Carry out the test as described in the "Electrophoresis" (Appendix 3.7) under both reducing and non-reducing conditions. Sample buffer (non-reducing conditions) Mix equal volumes of water and concentrated SDS-PAGE sample buffer. Sample buffer (reducing conditions) Mix equal volumes of water and concentrated SDS-PAGE sample buffer for reducing conditions containing 2-mercaptoethanol as the reducing agent. Reference solution (a) Solution of molecular weight markers suitable for calibrating SDS-polyacrylamide gels in the range of 14.4 to 94 kDa. Reference solution (b) Dilute Filgrastim RS with Sample buffer to obtain a concentration of 100 μg per mL. Test solution (a) Dilute Filgrastim Concentrated Solution with Sample buffer to obtain a concentration of 100 μg per mL. Test solution (b) To 200 μL of Test solution (a) add 200 μL of Sample buffer. Test solution (c) Dilute 200 μL of Test solution (b) to 1.0 mL with Sample buffer. Test solution (d) Dilute 200 μL of Test solution (c) to 1.0 mL with Sample buffer. Test solution (e) To 200 μL of Test solution (d) add 200 μL of Sample buffer. Electrophoresis System SAMPLE TREATMENT Boil for 5 minutes. GEL DIMENSIONS 1 mm thick. RESOLVING GEL 13 per cent acrylamide. APPLICATION 20 μL DETECTION Silver staining. System suitability SAMPLE Reference solution (a) the validation criteria are met;

No counterparts indexed yet — search this title across all datasets

View raw JSON-LD
{
  "@id": "https://www.openphar.org/data/tp/appendices/chromatography",
  "@type": "IndexedMonograph",
  "prefLabel": {
    "en": "Chromatography"
  },
  "edition": "TP-2011-Supplement-2024",
  "category": "appendices",
  "definition": {
    "en": "Mobile phase\n\nMOBILE PHASE A Trifluoroacetic acid, water, acetonitrile (0.5:950:50)\n\nMOBILE PHASE B Trifluoroacetic acid, water, acetonitrile (0.5:50:950)\n\nThe step gradient of mobile phases is as follows:\n\n<table border=\"1\"><tr><td>Time(Minutes)</td><td>Mobile Phase A(Per Cent V/V)</td><td>Mobile Phase B(Per Cent V/V)</td></tr><tr><td>0-8</td><td>97→94</td><td>3→6</td></tr><tr><td>8-25</td><td>94→66</td><td>6→34</td></tr><tr><td>25-40</td><td>66→10</td><td>34→90</td></tr><tr><td>40-45</td><td>10</td><td>90</td></tr></table>\n\n## Chromatographic system\n\nChromatographic system DETECTOR Ultraviolet light (215 nm). COLUMN A stainless steel column (10 cm $ \\times $ 2.1 mm), packed with octadecylsilyl bonded to porous\n\nsilica microparticles, compatible with 100 per cent aqueous mobile phases (5 $ \\mu\\mathrm{m} $ ) with a pore size of 20 nm, end-capped.\n\nCOLUMN TEMPERATURE 60°\n\nFLOW RATE 0.2 mL per minute\n\nSystem suitability\n\nSAMPLE Standard solution\n\nSuitability requirements The chromatogram obtained from the standard solution is qualitatively similar to the chromatogram of filgrastim digest supplied with Filgrastim RS.\n\nProcedure Inject a portion (about 10 mL) of Standard solution and Test solution into the chromatograph. Measure the areas of the principal peaks in each chromatogram. Identify the peaks due to filgrastim in the test solution using the retention time of filgrastim in the standard solution.\n\nResults The profile of the chromatogram obtained from the test solution corresponds to that of the chromatogram obtained from the standard solution.\n\nImpurities with molecular weights higher than that of filgrastim Carry out the test as described in the \"Size- Exclusion Chromatography\" (Appendix 3.6), use the normalization procedure.\n\nDiluent Dissolve 4.1 g of sodium acetate in 400 mL of water, adjust to pH 4.0 with acetic acid and dilute to 500 mL with water.\n\nStandard solution Dilute Filgrastim RS with Diluent to obtain a concentration of 0.4 mg per mL.\n\nTest solution"
  },
  "sections": [
    {
      "name": "Preamble",
      "text": "Mobile phase\nMOBILE PHASE A Trifluoroacetic acid, water, acetonitrile (0.5:950:50)\nMOBILE PHASE B Trifluoroacetic acid, water, acetonitrile (0.5:50:950)\nThe step gradient of mobile phases is as follows:\n<table border=\"1\"><tr><td>Time(Minutes)</td><td>Mobile Phase A(Per Cent V/V)</td><td>Mobile Phase B(Per Cent V/V)</td></tr><tr><td>0-8</td><td>97→94</td><td>3→6</td></tr><tr><td>8-25</td><td>94→66</td><td>6→34</td></tr><tr><td>25-40</td><td>66→10</td><td>34→90</td></tr><tr><td>40-45</td><td>10</td><td>90</td></tr></table>\nChromatographic system\nChromatographic system DETECTOR Ultraviolet light (215 nm). COLUMN A stainless steel column (10 cm \\times 2.1 mm), packed with octadecylsilyl bonded to porous\nsilica microparticles, compatible with 100 per cent aqueous mobile phases (5 μm ) with a pore size of 20 nm, end-capped.\nCOLUMN TEMPERATURE 60°\nFLOW RATE 0.2 mL per minute\nSystem suitability\nSAMPLE Standard solution\nSuitability requirements The chromatogram obtained from the standard solution is qualitatively similar to the chromatogram of filgrastim digest supplied with Filgrastim RS.\nProcedure Inject a portion (about 10 mL) of Standard solution and Test solution into the chromatograph. Measure the areas of the principal peaks in each chromatogram. Identify the peaks due to filgrastim in the test solution using the retention time of filgrastim in the standard solution.\nResults The profile of the chromatogram obtained from the test solution corresponds to that of the chromatogram obtained from the standard solution.\nImpurities with molecular weights higher than that of filgrastim Carry out the test as described in the \"Size- Exclusion Chromatography\" (Appendix 3.6), use the normalization procedure.\nDiluent Dissolve 4.1 g of sodium acetate in 400 mL of water, adjust to pH 4.0 with acetic acid and dilute to 500 mL with water.\nStandard solution Dilute Filgrastim RS with Diluent to obtain a concentration of 0.4 mg per mL.\nTest solution Dilute Filgrastim Concentrated Solution with Diluent to obtain a concentration of 0.4 mg per mL.\nResolution solution Mix a suitable portion of Standard solution for about 30 seconds using a vortex mixer.\nMobile phase Dissolve 7.9 g of ammonium hydrogen carbonate in 1000 mL of water and adjust to pH 7.0 with phosphoric acid; dilute to 2000 mL with water.\nChromatographic system\nDETECTOR Ultraviolet light (215 nm)\nCOLUMN A column (30 cm \\times 7.8 mm), packed with spherical silica (5 mm) with hydrophilic coating (suitable for fractionation of globular proteins in the relative molecular mass range of 10,000 to 500,000), equipped with a similarly packed guard column of a suitable size.\nCOLUMN TEMPERATURE 30°\nFLOW RATE 0.5 mL per minute\nSystem suitability\nSAMPLE Resolution solution\n(Note The relative retention with reference to the filgrastim monomer: aggregates, filgrastim oligomer 1, filgrastim oligomer 2, and filgrastim dimer are about 0.60,0.75,0.80,and 0.85,respectively,where the retention time for filgrastim monomer is about 19\nminutes.)\nSuitability requirements\nRETENTION TIME 17 to 20 minutes for filgrastim monomer.\nRESOLUTION Not less than 3 between filgrastim dimer and filgrastim monomer peaks.\nProcedure Separately inject equal volumes (about 20 μL ) of Standard solution and Test solution into the chromatograph, record the chromatograms and measure the responses for the major peaks.\nCalculation Calculate the percentage content of the dimer, oligomers and aggregates."
    },
    {
      "name": "Limits",
      "text": "- impurities with molecular weights higher than that of filgrastim, other than the dimer: not more than 0.5 per cent;\n- total of impurities with molecular weights higher than that of filgrastim: not more than 2 per cent.\nImpurities with molecular weights differing from that of filgrastim Carry out the test as described in the \"Electrophoresis\" (Appendix 3.7) under both reducing and non-reducing conditions.\nSample buffer (non-reducing conditions) Mix equal volumes of water and concentrated SDS-PAGE sample buffer.\nSample buffer (reducing conditions) Mix equal volumes of water and concentrated SDS-PAGE sample buffer for reducing conditions containing 2-mercaptoethanol as the reducing agent.\nReference solution (a) Solution of molecular weight markers suitable for calibrating SDS-polyacrylamide gels in the range of 14.4 to 94 kDa.\nReference solution (b) Dilute Filgrastim RS with Sample buffer to obtain a concentration of 100 μg per mL.\nTest solution (a) Dilute Filgrastim Concentrated Solution with Sample buffer to obtain a concentration of 100 μg per mL.\nTest solution (b) To 200 μL of Test solution (a) add 200 μL of Sample buffer.\nTest solution (c) Dilute 200 μL of Test solution (b) to 1.0 mL with Sample buffer.\nTest solution (d) Dilute 200 μL of Test solution (c) to 1.0 mL with Sample buffer.\nTest solution (e) To 200 μL of Test solution (d) add 200 μL of Sample buffer.\nElectrophoresis System\nSAMPLE TREATMENT Boil for 5 minutes.\nGEL DIMENSIONS 1 mm thick.\nRESOLVING GEL 13 per cent acrylamide.\nAPPLICATION 20 μL\nDETECTION Silver staining.\nSystem suitability\nSAMPLE Reference solution (a) the validation criteria are met;"
    }
  ]
}