1.10 WEIGHTS AND BALANCES
Definition
Pharmacopoeial tests and assays require the use of balance of capacity and sensitivity corresponding to the degree of accuracy sought. When substances must be "accurately weighed", the weighing shall be performed using a balance that is calibrated over the operating range and meets the requirements defined for repeatability and accuracy. For balances used for other applications, the balance repeatability and accuracy should be commensurate with the requirements for its use. ## Apparatus Analytical balances should possess adequate capacity and sensitivity. The analytical balance should be so constructed as to support its full capacity without developing undue stress and its sensitivity should not be altered by repeated weighing of the full-capacity load. The type of analytical balance having constant sensitivity over the whole capacity range is the constant-load, single-pan balance. It has a set of weights suspended from a counterpoised beam; in the process of weighing, these are removed from the beam by a manually operated mechanical device until equilibrium is reached. The analytical balance should be constructed in a proper housing with suitable openings to permit the placement of the material to be weighed and to exclude air currents. Desiccants may be placed inside the housing for the maintenance of a relatively dry atmosphere. Sets of calibrated weights used with balances that require manual placement of weights and sets of weights used to check the sensitivity of balances of another type should be kept in a case made of suitable material and properly lined. ## Placement of Balance The analytical balance should be placed upon a firm foundation that is as free from mechanical vibration as possible, preferably on an antivibration table of proper design. Alternatively, it may be placed on a concrete slab resting upon piers that are either sunk into the ground or connected to the construction elements of the building; or it may be placed upon a stout table or shelf protected by shock absorbers, such as cork mats or sheet rubber. The balance should also be protected from humidity and acid fumes, preferably by placing it in a separate room of the laboratory. It should not be placed in a current of air or in direct sunlight. The balance should be equipped with a levelling device and an indicator of proper position. Proper adjustment of levelling should be frequently checked. ## Checking of Sensitivity The sensitivity of the balance should be periodically checked by a qualified expert. ## Checking the Stability of the Equilibrium Position Before the balance is used, its equilibrium position without load should be checked several times. After each test, the balance has to be arrested. The equilibrium position of the balance under load should also be determined from time to time, for example, with one-tenth of the full load and with the full load. The difference between equilibrium positions found in two successive determinations made with equal loads should not exceed 0.1 mg for analytical balances and 0.0001 mg for analytical microbalances. ## Operation of the Balance When the balance is not in use, the balance beam and pan supports should be raised. The doors of the housing should always be kept closed. To release the balance, the beam and pans should be lowered very carefully. Objects to be weighed must be allowed to attain the temperature of the balance before weighing is started. The object to be weighed, as well as the weights, should always be placed on the pan as centrally as possible. During a weighing or on any occasion when objects are being added to or removed from the pans, both the beam arrests and the pan supports must be raised. Substances must be weighed in suitable containers such as beakers, weighing bottles, or crucibles. Liquids and volatile or hygroscopic solids must be weighed in tightly closed vessels, such as stoppered weighing bottles. No chemicals or objects that might injure the balance
Monograph text
Full text
Pharmacopoeial tests and assays require the use of balance of capacity and sensitivity corresponding to the degree of accuracy sought. When substances must be "accurately weighed", the weighing shall be performed using a balance that is calibrated over the operating range and meets the requirements defined for repeatability and accuracy. For balances used for other applications, the balance repeatability and accuracy should be commensurate with the requirements for its use. ## Apparatus Analytical balances should possess adequate capacity and sensitivity. The analytical balance should be so constructed as to support its full capacity without developing undue stress and its sensitivity should not be altered by repeated weighing of the full-capacity load. The type of analytical balance having constant sensitivity over the whole capacity range is the constant-load, single-pan balance. It has a set of weights suspended from a counterpoised beam; in the process of weighing, these are removed from the beam by a manually operated mechanical device until equilibrium is reached. The analytical balance should be constructed in a proper housing with suitable openings to permit the placement of the material to be weighed and to exclude air currents. Desiccants may be placed inside the housing for the maintenance of a relatively dry atmosphere. Sets of calibrated weights used with balances that require manual placement of weights and sets of weights used to check the sensitivity of balances of another type should be kept in a case made of suitable material and properly lined. ## Placement of Balance The analytical balance should be placed upon a firm foundation that is as free from mechanical vibration as possible, preferably on an antivibration table of proper design. Alternatively, it may be placed on a concrete slab resting upon piers that are either sunk into the ground or connected to the construction elements of the building; or it may be placed upon a stout table or shelf protected by shock absorbers, such as cork mats or sheet rubber. The balance should also be protected from humidity and acid fumes, preferably by placing it in a separate room of the laboratory. It should not be placed in a current of air or in direct sunlight. The balance should be equipped with a levelling device and an indicator of proper position. Proper adjustment of levelling should be frequently checked. ## Checking of Sensitivity The sensitivity of the balance should be periodically checked by a qualified expert. ## Checking the Stability of the Equilibrium Position Before the balance is used, its equilibrium position without load should be checked several times. After each test, the balance has to be arrested. The equilibrium position of the balance under load should also be determined from time to time, for example, with one-tenth of the full load and with the full load. The difference between equilibrium positions found in two successive determinations made with equal loads should not exceed 0.1 mg for analytical balances and 0.0001 mg for analytical microbalances. ## Operation of the Balance When the balance is not in use, the balance beam and pan supports should be raised. The doors of the housing should always be kept closed. To release the balance, the beam and pans should be lowered very carefully. Objects to be weighed must be allowed to attain the temperature of the balance before weighing is started. The object to be weighed, as well as the weights, should always be placed on the pan as centrally as possible. During a weighing or on any occasion when objects are being added to or removed from the pans, both the beam arrests and the pan supports must be raised. Substances must be weighed in suitable containers such as beakers, weighing bottles, or crucibles. Liquids and volatile or hygroscopic solids must be weighed in tightly closed vessels, such as stoppered weighing bottles. No chemicals or objects that might injure the balance pans should be placed directly upon them. When small quantities of a substance (for example, the sulfated ash) must be weighed in a large vessel and a fairly long period elapses between the two weighings, atmospheric pressure and temperature may alter sufficiently to affect the buoyancy and thus cause an appreciable error. In two-pan balances, this error may be eliminated by using another vessel of similar shape and weight for taring. The pans of the balance should be periodically lightly brushed with a camel-hair or similar brush to remove any dust that may have collected. The weights should be handled only by means of a pair of forceps, which should possess tips covered with suitable material. ## Calibration If necessary, turn on the power, and allow the balance to equilibrate for at least 1 hour before proceeding with the calibration. (Microbalances may require up to 24 hours to reach equilibrium.) ## REPEATABILITY Repeatability is assessed by weighing one test weight not less than 10 times. (Note The test weight must be within the balance's operating range, but the weight need not be calibrated. Because the standard deviation is virtually independent of sample mass within the balance's capacity, use of a small test weight, which may be difficult to handle, is not required.) Repeatability is satisfactory if twice the standard deviation of the weighed value, s, divided by the desired smallest net weight (i.e., smallest net weight that the users plan to use on that balance), does not exceed 0.10 per cent. The repeatability measurement establishes the smallest net amount of material that may be weighed on the balance in conformance with the 0.10 per cent limit. The minimum weight, M_{min} is described by the inequality M_{min} \geq 2000 s. For example, if s is found to be 0.00015, then M_{min} must be \geq 0.30000 g or 300.00 mg. If the standard deviation, s, obtained is less than 0.41d, where d is the scale interval of the balance, then the inequality bec
No counterparts indexed yet — search this title across all datasets
View raw JSON-LD
{
"@id": "https://www.openphar.org/data/tp/appendices/110-weights-and-balances",
"@type": "IndexedMonograph",
"prefLabel": {
"en": "1.10 WEIGHTS AND BALANCES"
},
"edition": "TP-2011-Vol-I-Part-1-Supplement-2020",
"category": "appendices",
"definition": {
"en": "Pharmacopoeial tests and assays require the use of balance of capacity and sensitivity corresponding to the degree of accuracy sought. When substances must be \"accurately weighed\", the weighing shall be performed using a balance that is calibrated over the operating range and meets the requirements defined for repeatability and accuracy. For balances used for other applications, the balance repeatability and accuracy should be commensurate with the requirements for its use.\n\n## Apparatus\n\nAnalytical balances should possess adequate capacity and sensitivity. The analytical balance should be so constructed as to support its full capacity without developing undue stress and its sensitivity should not be altered by repeated weighing of the full-capacity load. The type of analytical balance having constant sensitivity over the whole capacity range is the constant-load, single-pan balance. It has a set of weights suspended from a counterpoised beam; in the process of weighing, these are removed from the beam by a manually operated mechanical device until equilibrium is reached.\n\nThe analytical balance should be constructed in a proper housing with suitable openings to permit the placement of the material to be weighed and to exclude air currents. Desiccants may be placed inside the housing for the maintenance of a relatively dry atmosphere.\n\nSets of calibrated weights used with balances that require manual placement of weights and sets of weights used to check the sensitivity of balances of another type should be kept in a case made of suitable material and properly lined.\n\n## Placement of Balance\n\nThe analytical balance should be placed upon a firm foundation that is as free from mechanical vibration as possible, preferably on an antivibration table of proper design. Alternatively, it may be\n\nplaced on a concrete slab resting upon piers that are either sunk into the ground or connected to the construction elements of the building; or it may be placed upon a stout table or shelf protected by shock absorbers, such as cork mats or sheet rubber.\n\nThe balance should also be protected from humidity and acid fumes, preferably by placing it in a separate room of the laboratory. It should not be placed in a current of air or in direct sunlight.\n\nThe balance should be equipped with a levelling device and an indicator of proper position. Proper adjustment of levelling should be frequently checked.\n\n## Checking of Sensitivity\n\nThe sensitivity of the balance should be periodically checked by a qualified expert.\n\n## Checking the Stability of the Equilibrium Position\n\nBefore the balance is used, its equilibrium position without load should be checked several times. After each test, the balance has to be arrested.\n\nThe equilibrium position of the balance under load should also be determined from time to time, for example, with one-tenth of the full load and with the full load. The difference between equilibrium positions found in two successive determinations made with equal loads should not exceed 0.1 mg for analytical balances and 0.0001 mg for analytical microbalances.\n\n## Operation of the Balance\n\nWhen the balance is not in use, the balance beam and pan supports should be raised. The doors of the housing should always be kept closed.\n\nTo release the balance, the beam and pans should be lowered very carefully. Objects to be weighed must be allowed to attain the temperature of the balance before weighing is started. The object to be weighed, as well as the weights, should always be placed on the pan as centrally as possible. During a weighing or on any occasion when objects are being added to or removed from the pans, both the beam arrests and the pan supports must be raised. Substances must be weighed in suitable containers such as beakers, weighing bottles, or crucibles. Liquids and volatile or hygroscopic solids must be weighed in tightly closed vessels, such as stoppered weighing bottles. No chemicals or objects that might injure the balance"
},
"sections": [
{
"name": "Full text",
"text": "Pharmacopoeial tests and assays require the use of balance of capacity and sensitivity corresponding to the degree of accuracy sought. When substances must be \"accurately weighed\", the weighing shall be performed using a balance that is calibrated over the operating range and meets the requirements defined for repeatability and accuracy. For balances used for other applications, the balance repeatability and accuracy should be commensurate with the requirements for its use.\n\n## Apparatus\n\nAnalytical balances should possess adequate capacity and sensitivity. The analytical balance should be so constructed as to support its full capacity without developing undue stress and its sensitivity should not be altered by repeated weighing of the full-capacity load. The type of analytical balance having constant sensitivity over the whole capacity range is the constant-load, single-pan balance. It has a set of weights suspended from a counterpoised beam; in the process of weighing, these are removed from the beam by a manually operated mechanical device until equilibrium is reached.\n\nThe analytical balance should be constructed in a proper housing with suitable openings to permit the placement of the material to be weighed and to exclude air currents. Desiccants may be placed inside the housing for the maintenance of a relatively dry atmosphere.\n\nSets of calibrated weights used with balances that require manual placement of weights and sets of weights used to check the sensitivity of balances of another type should be kept in a case made of suitable material and properly lined.\n\n## Placement of Balance\n\nThe analytical balance should be placed upon a firm foundation that is as free from mechanical vibration as possible, preferably on an antivibration table of proper design. Alternatively, it may be\n\nplaced on a concrete slab resting upon piers that are either sunk into the ground or connected to the construction elements of the building; or it may be placed upon a stout table or shelf protected by shock absorbers, such as cork mats or sheet rubber.\n\nThe balance should also be protected from humidity and acid fumes, preferably by placing it in a separate room of the laboratory. It should not be placed in a current of air or in direct sunlight.\n\nThe balance should be equipped with a levelling device and an indicator of proper position. Proper adjustment of levelling should be frequently checked.\n\n## Checking of Sensitivity\n\nThe sensitivity of the balance should be periodically checked by a qualified expert.\n\n## Checking the Stability of the Equilibrium Position\n\nBefore the balance is used, its equilibrium position without load should be checked several times. After each test, the balance has to be arrested.\n\nThe equilibrium position of the balance under load should also be determined from time to time, for example, with one-tenth of the full load and with the full load. The difference between equilibrium positions found in two successive determinations made with equal loads should not exceed 0.1 mg for analytical balances and 0.0001 mg for analytical microbalances.\n\n## Operation of the Balance\n\nWhen the balance is not in use, the balance beam and pan supports should be raised. The doors of the housing should always be kept closed.\n\nTo release the balance, the beam and pans should be lowered very carefully. Objects to be weighed must be allowed to attain the temperature of the balance before weighing is started. The object to be weighed, as well as the weights, should always be placed on the pan as centrally as possible. During a weighing or on any occasion when objects are being added to or removed from the pans, both the beam arrests and the pan supports must be raised. Substances must be weighed in suitable containers such as beakers, weighing bottles, or crucibles. Liquids and volatile or hygroscopic solids must be weighed in tightly closed vessels, such as stoppered weighing bottles. No chemicals or objects that might injure the balance pans should be placed directly upon them.\n\nWhen small quantities of a substance (for example, the sulfated ash) must be weighed in a large vessel and a fairly long period elapses between the two\n\nweighings, atmospheric pressure and temperature may alter sufficiently to affect the buoyancy and thus cause an appreciable error. In two-pan balances, this error may be eliminated by using another vessel of similar shape and weight for taring.\n\nThe pans of the balance should be periodically lightly brushed with a camel-hair or similar brush to remove any dust that may have collected.\n\nThe weights should be handled only by means of a pair of forceps, which should possess tips covered with suitable material.\n\n## Calibration\n\nIf necessary, turn on the power, and allow the balance to equilibrate for at least 1 hour before proceeding with the calibration. (Microbalances may require up to 24 hours to reach equilibrium.)\n\n## REPEATABILITY\n\nRepeatability is assessed by weighing one test weight not less than 10 times. (Note The test weight must be within the balance's operating range, but the weight need not be calibrated. Because the standard deviation is virtually independent of sample mass within the balance's capacity, use of a small test weight, which may be difficult to handle, is not required.)\n\nRepeatability is satisfactory if twice the standard deviation of the weighed value, s, divided by the desired smallest net weight (i.e., smallest net weight that the users plan to use on that balance), does not exceed 0.10 per cent. The repeatability measurement establishes the smallest net amount of material that may be weighed on the balance in conformance with the 0.10 per cent limit. The minimum weight, M_{min} is described by the inequality M_{min} \\geq 2000 s. For example, if s is found to be 0.00015, then M_{min} must be \\geq 0.30000 g or 300.00 mg. If the standard deviation, s, obtained is less than 0.41d, where d is the scale interval of the balance, then the inequality bec"
}
]
}